Check analog connections using nominal potentiometer values #100

Open
adeguet1 opened this issue Aug 9, 2018 · 0 comments
Open

Check analog connections using nominal potentiometer values #100

adeguet1 opened this issue Aug 9, 2018 · 0 comments
Assignees

Comments

@adeguet1
Copy link
Contributor

adeguet1 commented Aug 9, 2018

We can use the potentiometer values to help identify the arm connected or at least make sure the values read make sense for the arm we're supposed to control. E.g. for ECM, only the first 4 values should be meaningful and should be within a specific range. For a PSM, the last pot is not used. For a MTM, all pots are used. This can be used to detect:

  • which arm is connected
  • faulty potentiometer
  • overall connections issues (SCSI between QLA and dMIB) or Cannon ITT connection

See also #99